Respiration Frequency Monitor: RM-80
The RM-80 is a stand-alone system comprising a pressure sensor, chamber and a fresh air delivery system. The instrument displays respiration rate in breaths per minute. An apnea detection system signals an alarm condition if no respiration is detected for 10 seconds. The RM-80 provides connections that supply signals related to the monitored respiratory parameters. An analog voltage is available that is proportional to the pressure changes within the chamber. In addition, an analog voltage is available that is proportional to the rate of respiration. These signals may be presented to strip chart recorders or data acquisition systems. Other outputs are provided that indicate apnea and respiratory activity in the form of open collector digital outputs. These event signals may be accumulated by employing any of Columbus Instruments' Computer Counter Interface products. Features / SpecificationsHigh SensitivitySimple stand-alone operation with large display of respiration frequency in breaths per minute Apnea sensing and alarm Adaptable to animals of all sizes Measures under continuous air flow Monitoring Range: 20 to 720 breaths per minute Sensitivity: Capable of monitoring a 20 gram mouse in a 3 liter chamber Accuracy: 1%, ±1 digit; 60 to 600 breaths per Minute Outputs: Analog pressure wave form and apnea alarm Analog rate voltage: 10mv/bpm
Ordering Information1500 Respiration Frequency Monitor RM-801510 Respiration Frequency Monitor RM-80 Interface to PC 1511-1 Respiration Frequency Monitor RM-80 Chamber, 1 liter 1511-2 Respiration Frequency Monitor RM-80 Chamber, 2.5 liter 1520 Respiration Freq Monitor RS-232 Wave Form Display and Interface |